The S2K controls the following functions:
                    
                    1. A 3-way solenoid valve: A DIP switch is used to set the 
                    amount of time the sample is to be exposed to the sensor. 
                    The rest of the time the sensor is exposed to clean air.
                    
                    2. An external pump control timer which sets the on-off time 
                    of the pump. The purpose of the external pump is to transport 
                    the sample in larger quantities which shortens the transport 
                    time. Additionally, it is easier to draw a controlled amount 
                    of sample to be conditioned before exposing it to the sensor. 
                    The timer turns the pump on only as needed which prolongs 
                    the pump life and reduces pump maintenance.
                    
                    3. A smaller, internal pump is on constantly to flow either 
                    sample or clean air to the sensor. There are three timers 
                    required to implement the complete cycle as shown in Fig. 
                    2 below. 
                    
                    
                    
                    1. External pump delay: This is the time delay from the beginning 
                    of the cycle before the external pump is activated. During 
                    this time, the external pump is off. The timer can be set 
                    from 0-255 minutes, with a resolution of 1 minute.
                    
                    2. External Pump Lead Time: This is the amount of time the 
                    external pump is "ON", before the solenoid valve 
                    is switched to let the sample into the sensor chamber. The 
                    timer can be set from 0-2550 seconds, with a resolution of 
                    10 seconds.
                    
                    3. This is the amount of time the solenoid valve is activated 
                    to allow sample flow into the sensor chamber. The timer can 
                    be set from 0-2550 seconds, with a resolution of 10 seconds.
                    
Each timer is set by an eight position DIP switch with 
                      the following count values:
                      1 + 2 + 4 + 8 + 16 + 32 + 64 + 128 = 255. This means that 
                      position one of the DIP switch has a value of "1" 
                      while position 8 has a value of "128". The time 
                      which is set for the timer is calculated by taking the count 
                      value and multiplying it by the resolution of the timer. 
                      A time of zero is achieved by having all of the switches 
                      in the open position.
                    There is a micro-switch which activates a relay contact 
                      if there is a failure of the flow into the sensor chamber. 
                      A calibration switch with an LED indicator allows the user 
                      to perform standard calibration functions on the sensors. 
                      The S2K housing is designed to be used with standard IST 
                      4-20 IQ sensor modules but other sensor modules (such as 
                      the SM95) can be used also.
                    The S2K sampling system can be configured with a variety 
                      of options. Fig. 3 illustrates the two-stage sampling system. 
                      Figure 4 shows a complete system with a sample diluting 
                      assembly. IST will assemble the system to meet application 
                      requirements.
                      
                    
                    
                      
                    
                    
                    
                       
                        System Specifications: 
                           
                           | 
                         | 
                      
                       
                        Enclosure: 
                           
                         | 
                        NEMA 4X. | 
                      
                       
                        Dimensions: 
                           
                         | 
                        345(H) x 292(W) x 127(D) mm; (13.6"H 
                          x 11.5"W x 5.0"D). | 
                      
                       
                        Weight: 
                           
                         | 
                        5.5 Kg (12 Lbs). | 
                      
                       
                        Timer controller function: 
                           
                         | 
                        See description above. | 
                      
                       
                        | Relay: | 
                        SPST Normally Open, 5 amp resistive. 220 
                          VAC max. 1/8 HP motor. 
                          Three relays are provided for the following: 
                          1. No flow into sensor chamber. 
                          2. External pump activation Relay.  
                          3. Solenoid valve. 
                         | 
                      
                       
                        Internal Pump: 
                           
                         | 
                        Oscillating AC Diaphragm. Replaceable. 
                         | 
                      
                       
                        Flow Meter: 
                           
                         | 
                        Dwyer Instrument, Inc. RMA-13-SSV. Ranges 
                          as specified. | 
                      
                       
                        Power: 
                           
                         | 
                        110/220 VAC, 50/60 Hz. | 
                      
                       
                        Temperature Range: 
                           
                         | 
                        -20°C to 50°C. | 
                      
                       
                        | Humidity:  | 
                        Non-condensing. |